Heat studies performance exami

RZ series high temperature thermogravimeter

The instrument is used to measure the change of mass of samples heated in different atmospheres. The equipment is designed an

The instrument is used to measure the change of mass of samples heated in different atmospheres. The equipment is designed and manufactured according to the requirements of GB / T 17732 "test methods for dense shaped carbon containing refractories" and the research of high temperature nitriding process of refractories. IPC industrial computer is used for automatic control. The interface is friendly, the temperature control precision is high (expert fuzzy control algorithm is applied), and the test results are recorded automatically. The curves of temperature time, quality time and quality temperature can be drawn. The original data can be converted into excel storage format, which can be read, queried and printed at any time.

RZ系列高温热重仪(图1)

 型号/参数 Model/specificationRZ17-01RZ17-02RZ17-03
 使用温度 Working temperature(℃)1650
 升温速率 Heating-up rate(℃/min)0~8
 样品最大质量 Max.mass of sample(g)50200500~1500
 测量精度 Measuring accuracy(g)0.00010.0010.01
 保留温度精度 Temperature accuracy in soaking(℃)±2
 炉膛尺寸 Hearth size(mm)200×150×150
 炉膛气氛 Atmosphere of the furnaceAir,N₂,Ar
 炉膛压力 Air pressure of hearth(MPa)-0.095~0.06
 主机尺寸 Size of mainframe(mm)800×610×1700
 电源 Power supply380V 50Hz 12kW
 重量 Weight(kg)800